Vasona Issues and Updates | ||
This is a log of the recent changes and current issues. | ||
Vasona Change Log | ||
This is a log of the recent changes to Vasona | ||
Version | Changes | Status/Conditions |
6.0098 |
Marker error fix for the MXA |
|
6.0097 |
Correctly store fwd
calculations during the fixed fowarrd power method |
|
6.0095 |
Added further Tower/turntable investigation data functionality |
|
6.0094 |
Tower/Turntable data stored as
a real number, not integer |
|
6.0093 |
Improved tracking of tower/turntable data |
|
6.0092 |
Smart fix for reversing positions |
|
6.0091 |
Additional checking that the
tower/turntable has got to their correct positions |
|
6.0089 |
New menu for tower/turntable
audit |
|
6.0088 |
Allow FSEK to work correctly
with new default IEEE settings. |
|
6.0087 |
Further development of the
driver of the ESW |
|
6.0086 |
Support Win10 network
functionality |
|
6.0084 |
Improved
messaging on the status of the pre-amp |
|
6.0083 |
Further
changes on how the switch driver works.
Improvements
to the turntable accuracy process. Additional
messaging during trace average function in combination with External
filtering. |
|
6.0081 |
Further
Improvements of the trace average function. |
|
6.0078 |
Improvements
of the trace average function and when elements are sent to the spectrum
analyser. |
|
6.0077 |
Change
how the switch driver works for various measurements. |
|
6.0076 |
Further
development of the TDemi analyser driver. |
|
6.0075 |
Further development of the TDemi analyser driver. Stopped
the error message about setting the Antenna height. |
|
6.0074 |
Stop
a problem where the maximization window constantly communicated to the Sunol
turntables. |
|
6.0073 |
Win10
fix to allow the table to have more than 6 points. |
|
6.0072 |
Improved
messaging when an external preselector is used. |
|
6.0071 |
Bug fix to ensure all analyser drivers to work correctly with dual inputs. Ensure
that when the template has both a signal generator and a spectrum analyser,
that the frequency range of the signal generator is not considered in all
cases. |
|
6.0069 |
Force
measurement to be ‘None’ when the template is not measuring a second trace
during FFT/Time Domain measurements. |
|
6.0068 |
Began
development of the error trap process.
|
|
6.0067 |
Support
more dongles. |
|
6.0066 |
Continued
development of driver for AR FM4000 |
|
6.0065 |
Improvements
in SmartMotor driver. |
|
6.0063 |
Limit
Svswr to every 50MHz steps (if needed) |
|
6.0062 |
Added
formal test attenuation mode. |
|
6.0061 |
Error
trapping during formal and debug scans |
|
6.0059 |
Improvement
measurement accuracy of tower / turntable positioning. |
|
6.0058 |
Added
a ms holdtime code function |
|
6.0057 |
Added
broadband fine tune method and a narrowband option during the process
rather than having to set them individually. |
|
6.0056 |
Minor
bug fix for modulation detection response. |
|
6.0055 |
Added
Method to detect if the AM modulation is not being applied correctly. |
|
6.0054 |
Insert
loss now has dedicated functionality |
|
6.0053 |
Allow
setting of the which active slot the emCenter controller is using. |
|
6.0052 |
Show
reset count of ISOtropic field probe in test results window |
|
6.0051 |
Modifications
to allow measurements above 100 GHz Development of drivers for ISOtropic field probes using IEEE. |
IsoTropic field probe control still in development |
6.0049 |
Added
option to allow further control of amplifier switch band timing using the
relevant driver. MXA, improvements to analsyer/rx switching process. Allow better storing of equipment type in the IEEE control screen. |
|
6.0048 | HP8546A, allow measurements above 1GHz using second port. | |
6.0047 | Added a copy of all files to the clipboard after automatic network storage. | |
6.0046 |
Added
ability to select a different location for the files which will be copied using
the auto network storage function. Stored name of computer in the results in the emi file. Improved during auto network storage function. |
|
6.0045 |
HP8546A
fix to stop the use of the *WAI command, needed specifically for the ESR/ESW
only. Added a specific '[end of file]' flag at the end of the emi file. Simplified process for recall of processed tower data. Improvements to timing of ISOtropic field probe measurments. |
|
6.0044 |
Added
a specific hold time for later Keysight/Agilent power meters. Removed storage of template request when the template is read only. Corrected analyser reset when using spectrum analyser to monitor forward power during conducted immunity calibration. Corrected antenna factor reporting during radio performance measurements. |
Still working on improvements to template request messaging. |
6.0043 | Provide results of network analyser results when the trace is negative. | |
6.0042 | Added driver for Keysight 19xx series. | |
6.0041 |
Added
additional delays to driver for HD tower/turntabbles. Additional menu items for narrowband tower/turntable maximization. |
|
6.0039 | Added option for the exclusion of the test details from the HTML output. | |
6.0038 | Bug fixes for the Frankonia tower/turntable drivers. | |
6.0037 | ||
6.0036 | ||
6.0035 |
Ability
to define the sweep time and number of points during a formal measurement. Bug fix in sweep table function to allow data to be copied and pasted. Bug fix in R&S ESCI, when using more than one specification limit Support for the most up to date Sunol tower/turntable controller |
|
6.0033 | Ability to record the forward power during a test. | |
6.0032 | Allow control of R&S ESW pre-amplifier. | |
Driver for Agilent N1913. | ||
6.0031 | Update to storage of remote radio templates. | |
Bug fix for Agilent PXA setting marker to center freqeuncy. | ||
6.0029 | Driver for EMCenter (ETS Lindgren) | |
Completed functionality to simple copy results files to the network. | ||
6.0028 | Big fix to allow more analyser to use 'rms' detectors. | |
6.0027 | When toggling analyser inputs, and apply change from the sweep details window, a preview function is operated. This will toggle the input from 1 to 2, or 2 to 1. | |
Completion of R&S ESR driver, including time domain mode | ||
For supported network drives, data is automatically copied. Plus function to copy files to the additional network folder. | ||
Bug fix on R&S ESCS receive sweep mode. | ||
Bug fix in maximisation which performs a fine tune, when requested no fine tune. | ||
Short cut to toggle to Res/Vid/Span function when in RX control mode. | ||
R&S ESCI, ESU, ESR. Marker to peak center function working fully. | ||
R&S ESCI, storage of step when toggling between Rx Control and Anayser control. Plus toggling to Rx mode more consistent. | ||
Ability to open files with more than 32000 data points in the scan data. | ||
Short cut function to plot both traces without having to open sweep details window. | ||
6.0025 | Improved measurement of narrowband signals during formal measurements. This sets the video bandwidth low so that it removes all the noise and azimuth / height settings are dramatically improved. | |
Further RX sweep and time domain developments | ||
6.0024 | Complete development of the Narda EP200 probe driver | |
Bug fix to set the level to dBuV for calibration data file creation. | ||
6.0023 | Continuing development of the Narda EP200 probe driver | |
Ability to paste data into the tower/turntable results | ||
6.0022 | Initial development of the Narda EP200 probe driver | |
6.0021 | Ability to set set size base around bandwidth (for rx stepped measurements) | |
Development of the R&S ESH3/ESVP driver | ||
6.0019 | Improved setting of the marker on the Agilent MXE. | |
6.0018 | Ability for the Agilent 74xx to perform rms measurements | |
6.0017 | Ability to reset table scan function back to those within the file | |
Improvements in the Narda probe drivers | ||
6.0016 |
Scan
table functionality. This is a significant update, needed to consolidate templates and allows the control of Attenuation Reference Level Resolution Bandwidth (ResBW) Video Bandwidth (VidBW) Bandwidth for formal tests (for some analyser the VidBW also has an impact on the measurement). These are settable based upon frequency range. This is needed for radio and marine testing. Plotting Resolution Bw plotted (the one used during the sweep) Fix to stop 40GHz, present as 39999.999GHz. ESCI Control of the IF analysis sweep screen during formal testing to stop 'relay switching' GR1089 Allow users to store specification limits for GR1089 (specifically sectoin 10). The use to be an admin only function. EN50083-2 During Immunity testing, the ability to apply two CW signals based upon the requirements of EN50083-2. |
Because
of the nature of the change we recommend significant debug testing of templates
prior to employment into a production environment. Another recent discussion point, LIMITs defined in dBpW are not always correctly displayed. Some analysers will apply a 17 dB offset if dBpW has been selected. Make sure you apply the correct factor. NOTE: the standard correction factor installed in the ECSI may not give you correct results. |
Bug
fixes Vasona would crash opening sweep details window in existing files. Vasona would still do a fine tune during conducted emission measurements, even though the signal was defined as 'no fine tune'. When antenna is fixed, vertical polarity was not available. Improvements on the application of offsets for immunity testing. Vasona will not apply the correct offset if the limits are defined in dBuA. |
||
6.0015 |
Driver
for R&S NRVD Improvements on user interface for GR1089 calculations. |
|
6.0014 |
Conducted
Immunity Allow the use of spectrum analysers for recording levels during conducted immunity testing. Provided a 'min hold' function to ensure levels are high enough during GR1089 section 10 calibration. Modification in section 10 calculation so that Amps are used, rather than mA. Tower/Table Control Improvements in controlling the 'max full' tower/turntables. |
|
Bug
fixes Formal measurements using the RMS detector use correct settings. |
||
6.0013 |
Started
development of the driver for the R&S FSW Added Fluke as an manufacturer, further improved driver for Smart field monitor. Error trap to make sure the correct second device is attached for CI calibration. Further modification of Error messages. Allow AM mod frequency, other than 1000. |
Prescan functionality of ESR and FSW should work. |
6.0012 | Added delay to find peak to allow slower analyser to be used for conducted immunity calibration. | |
6.0011 | Added SMart field monitor probe. | This uses some custom code, to use revision after this, may require a complete install. Noting that the NI GPIB drivers will need to re-installed afterwards. |
6.0009 | Allow
the selection of the second port for ESU/MXE. Started development of the driver for the R&S ESR Started development to allow the control of the number of points used during prescan. |
|
6.0008 | Completed development of the R&S automated LISN control. | |
6.0007 | Allow control of very early Sunol tower/table controller | |
6.0006 | During continuous capture, a second dummy window would be created. Issue resolved. | |
Stop crashing when trying to open the RVC window. | ||
GR1089 Section 10 calculations, removed limitation on calculations to 200A | ||
Bug
fixing in Ethernet mode Fixed formal ESCI/ESU and Maxfull control issues Soak tested data capture |
Beta | |
6.0005 | Bug fixing in Ethernet mode | Beta |
6.0004 | Added control of device over Ethernet, 15% faster data capture | Beta release |
6.0003 | Added automatic control for LISNs (under prescan function) | Beta release |
6.0002 | Update driver for R&S FSP analyser [1.4] | Removed error messages |
Improved storage of object files (archive) | ||
6.0001 | Added additional driver for MaxFull tower/turntables | |
5.0099 | Bug fix opening some files. | |
Ability to change the size of text on HTML files | ||
When merging RI calibration files, the overall results could now be saved to an EMI file. | ||
5.0098 | NSA, updated process so that the figures are always saved. | Prior to this, the figures were only saved after opening the NSA Results window and then hitting save. |
Added an additional Save option in NSA Window | ||
Added
additional 'Cancel' message when the user cancelled the current test. Additional information (in messages) about what has just been tested. |
||
BETA,
save template function (from within the
results section).. |
Very useful when generating new templates, not need to toggle back/forward | |
8566,
update specification limit drawing so that the bright spot is removed at the end
of the line. Autodraw of limits when changing the span Created driver for 8568B. |
||
5.0097 | Allow the use of the Sunol SC110V tower/table controller | Fix from 5.0095 |
R&S ESHS10 Driver [1.3] | Cleared missing element | |
5.0096 |
R&S ESI, fix when using receiver control, Vasona hangs when measuring Qp, Av, Pk or RMS measurements. |
This does not occur in analyser mode. |
Further checks to ensure Vasona is controlling the R&S ESI/ESU/ESL in rx mode. | ||
Driver complete for R&S ESL | ||
5.0095 | For Svswr measurements, plots can be frequency limited. | Use the apply function in the sweep details window and the copy 'Full graph' to clipboard. |
Ability to connect to a Sunol 110 series firmware | Note, because of the changes in the firmware, vasona still does not control the towers/turntables correctly using this firmware version. | |
Added better control of timing using older HP analysers where Vasona forces the sweep to occur in the preselector bands. (above 1GHz) | This was initially done for the HP856x series. If other analysers have the problem, issues can be fixed in the analyser driver. | |
R&S ESL driver in development | This is based on a ESI. | |
RI
calibration fixes. 1. Forward power/Driver table selection is now automatic at the completion of RI calibration. 2. Resaving of drive tables without forward power does not create an error. 3. Only the save-as function is available (to avoid confusion) 4. RI Calibration window, is now auto-sized. |
||
5.0094 | Add diagram to better illustrate the setup for Svswr measurements. | |
5.0093 | Improved storage of files on the network. | Results are not longer limited to the default directory. |
Improved visibility of the mapped networked drivers and when the are connected. | ||
Improved PING functionality | ||
5.0092 | Linearity check for conducted immunity calibration, data is stored in the emi file. | |
Corrected functionality of limited dongles. | ||
Improved saving of calibration factors to the 'cal factors' directory. | ||
Agilent MXE, PXA, additional dwell included when setting marker frequency. | The marker functions of these analysers do not work correctly and hence Vasona has to put additional delays to ensure that the display is updated. | |
When
measuring injected current 1. improved initialization of the analyser used to do the measurements. 2. The sweep time function works, rather than auto. |
||
Using library frequencies, defaults are now not selected on the second run of the test. | ||
HMTL output, format fixed so that the image is always available. | Prior to this only IE gave the correct output. | |
RTF Output has been investigated. | Note, code has not been implemented but we have solved any technical difficulties associated with outputting data in RTF format. | |
RI Calibration, ability to merge separate result together. | This functionality should be carefully checked prior to use. | |
Ability to see the composite specification from the admin menu. | ||
For R&S ESU and ESCI, when the limit is defined in dBuA, results have a 34 dB offset if Vasona does the correction. | ||
For
newer Agilent analysers, second limit line is always deleted. This
includes, MXE 740x, PSA. |
||
Add log steps | Need for RVC calibration. | |
Start the development of Common storage filenames. | This functionality will improved how common templates are stored and used. This will take several iterations fix correctly. | |
5.0091 | RVC calibration calculations are now built-in to Vasona. | |
Corrected functionality of limited dongles and the display of which options they control. | ||
Improved saving of calibration factors to the 'cal factors' directory. | ||
Additional checking to make sure that the ISOtropic field probes port is available prior to radiated immunity calibration. | ||
Vasona Issues | ||
This is a list of none issues and work items. | ||
Version | Issue | Work arounds |
When operating different templates with a PSA and an ESCI, Vasona can stop prior to formal testing. | Reboot Vasona between tests. | |
ESR driver needed | Debug scans work in Version 5.0009. | |
RVC Testing functionality | In development | |
Scan receiver functionality (new) | In development |
The other details.... |
HDR.Expert, EMiSoft © 1997-2019 [Trademarks mentioned herein are the property of their respective owners.] |